Portable pH meter AZ 8690
- Stock: Out of Stock
- Model: AZI-8690
- EAN: 5907222598726
- Weight: 75g
- Dimensions: 42mm x 37mm x 149mm
AZ 8690 is a handy meter for measuring the pH value and temperature of the tested liquid. It is characterized by a high measuring range (0 ~ 14.00 pH) and a resolution of indications up to hundredths of pH.
Placing the display at a 45° angle (patent protected solution) makes it easier to read the indications. Possibility of replacing the electrode minimizes the operating costs of the device. Measurement results are displayed on a two-field LCD panel with an accuracy of ±0.05pH and ±0.6°C / ±0.8°C for temperature.
Meter has an automatic three-point calibration function that can be carried out using standard buffers (pH: 4.0 / 7.0 / 10.0), an automatic temperature compensation function and HOLD function that allows you to freeze the indications on the display.
AZ 8690 is powered by 2 x batteries, type CR2032. It has an automatic power cut-off function after 20 minutes of inactivity, with an option to deactivate it and a low battery indication.
APPLICATION
AZ 8690 will be perfect as a measuring device in school chemical laboratories, photo laboratories, for testing the pH reaction in aquariums, swimming pools, gardens, etc.
ADVANTAGES
- Two-field LCD display
- High measurement accuracy
- ATC function - automatic temperature compensation
- HOLD function - stopping the indications
- Temperature measurement units can be changed: °C / °F
- Waterproof housing IP 67
- Low battery indication
- Replaceable electrode
- Automatic power cut-off after 20 minutes, with option to deactivate
- Small dimensions and weight
Product information | |
Display | LCD 12 x 19 mm, dual-pole |
Probe calibration | Automatic, three-point - standard pH buffers: 4.0 / 7.0 / 10.0 |
Probe life time | > 6 months (with proper use and maintenance) |
Additional features | HOLD - freezing indications on the display ATC - automatic temperature compensation* * Enable measurement in various environmental conditions without the need to recalibrate the meter |
Additionally included | 2 x batteries, type CR2032 Material pouch User manual |
pH measurement | |
Measurement range | 0 ~ 14.00 pH |
Measurement accuracy | ±0.05 pH * Accuracy may decrease with use. It is recommended to calibrate the device regularly. |
Measurement resolution | 0.01 |
Temperature measurement | |
Measurement range | 0 ~ 50°C |
Accuracy of measurement | ±0.6°C for 20 ~ 30°C ±0.8°C or other ranges * Accuracy may decrease with use. It is recommended to calibrate the device regularly. |
Measurement resolution | 0.1 |
Temperature coefficient | Automatic temperature compensation for the entire measuring range |
Units of measurement | °C / °F |
Power | |
Power source | 2 x batteries, type CR2032 |
Energy saving | Automatic power off after 20 minutes of inactivity, with deactivation option |
Battery discharge indication | Yes |
Working and storage conditions | |
Working conditions | Temperature: 0 ~ 50°C Humidity: up to 80% RH, non-condensing |
Storage conditions | Temperature: -10 ~ 50°C - without batteries & probe Humidity: up to 90% RH non-condensing |
Quality and safety | |
Certificates and approvals | CE mark |