Previous
Low temperature data logger AZ 88165
Extremely low temperature data logger with probe AZ 88264
- Stock: Out of Stock
- Model: AZI-88264
- Weight: 50g
- Dimensions: 45mm x 17mm x 110mm
AZ 88264 is a data logger for extremely low temperatures with a USB output for data download and an advanced external probe.
The probe has PT1000 temperature sensor and has been specially designed to measure extremely low temperatures. It enables temperature registration in the range of -100 ~ 120°C. The whole thing is mounted on a silicone cable with a length of approx. 200 cm. The results are stored in the memory with a total capacity of 48,000 measurement points.
Programming the recorder (e.g. changing the interface language, setting the sampling frequency, alarm thresholds, delaying the recorder activation, units for temperature, etc.) and reading the results is done using a computer without the need to install additional software. After connecting the recorder to a computer, a measurement report is automatically generated in the form of a PDF file, which contains the compilation of measurement data. Thanks to this, the user has the opportunity to quickly assess the conditions in which the measurements took place.
AZ 88264 is powered by 2 x 3V batteries, type CR2032. The batteries last for about 3 months of uninterrupted operation of the device. Each recorder has a unique serial number that enables unambiguous identification, and also has the option of entering a username and a password.
APPLICATION
AZ 88264 is suitable for monitoring and controlling temperature in industrial refrigerators and cold rooms, refrigeration machines, containers for storing and transporting food and drugs, as well as in warehouses, laboratories, cellars, etc.
ADVANTAGES
- Very wide measuring range (-100 ~ 120°C)
- External probe on a 200 cm long silicone cable
- Built-in PT1000 sensor for detecting extremely low temperatures
- Operation of the recorder without the need to install additional software
- Memory of total capacity of 48,000 measurement points
- Sampling time set in the range: 30 sec. ~ 120min.
- Possibility to set alarm thresholds and the delay time of the recorder activation
- Signaling of exceeding the set alarm thresholds
- LED diodes - green and red informing about the operating mode
- MARK button - marker for recording important events
- Automatic generation of data reports in PDF format
- You can replace the battery yourself
- Factory calibrated device
- Compact dimensions, easy operation
Product information | |
Product type | Multi-use |
Memory | 48'000 samples for temperature (24'000 int. device & 24'000 etx. probe) |
Alarm type | Single event / collective alarm / no alarm |
Alarm thresholds | Programmable at range: -30 ~ 70°C |
Alarm delay | 0, 5, 30, 45, 60, 90, 120 minutes |
Activation delay | 0, 5, 30, 45, 60, 90, 120 minutes |
Sampling frequency | 30 sec; 5, 10, 30, 60, 90, 120 min. |
Buttons | Start / Stop MARK - marker of important events |
LED lights | REC, OK, LOW/HI alarm |
Sockets, connectors, ports | USB 2.0 - for downloading data to a computer |
Additional features | Probe with PT1000 sensor for detecting extremely low temperatures Option to set a password to protect against opening |
Additionally included | 2 x 3V battery, type CR2032 External probe with PT1000 sensor User manual |
Temperature measurement | |
Measurement range | Datalogger: -30 ~ 70°C (-22 ~ 158°F) Ext. probe: -100 ~ 120°C (-148 ~ 248°F) |
Accuracy of measurement | Datalogger: ±0.5°C Ext. probe*: ±0.5°C * Logger placed at temperature: 10 ~ 40°C. ** Accuracy may decrease with use. It is recommended to calibrate the device regularly. |
Measurement resolution | 0.1 |
Units of measurement | °C / °F |
Power | |
Power source | 2 x 3V battery, type CR2032 (can be replaced) |
Battery life | ±3 months from the moment of activation* * This is the maximum battery life. Battery life can be influenced by e.g. high activation frequency alarm, exposure of the device to temperatures outside the operating range specified in the specification, etc. |
Battery discharge indication | Yes |
Working and storage conditions | |
Working conditions | Temperature: -70 ~ 80°C (device); -100 ~ 120°C (ext. probe) Humidity: up to 90% RH, non-condensing * Room temperature ±25°C - when uploading data to a PC |
Storage conditions | Temperature: -40 ~ 85°C Humidity: up to 90% RH non-condensing |
Quality and safety | |
Certificates and approvals | CE mark |