Thermoanemometer AZ 8911
- Stock: Out of Stock
- Model: AZI-8911
- EAN: 5907222598832
- Weight: 170g
- Dimensions: 70mm x 33mm x 175mm
Available Options
AZ 8911 thermoanemometer is a professional meter used to measure the speed and efficiency of air flow and to measure relative humidity, current temperature with an additional function of calculating the dew point and evaporation temperature.
Sapphire rotor bearing ensures a very accurate measurement even at low air velocities. User can change the temperature reading units (°C or °F) and the flow velocity (m/s or ft/min).
Device can operate in the following measurement modes: dynamic, minimum / maximum and average reading. HOLD function makes it possible to freeze the indications with a quick response. Vane probe is connected to the meter with a 1.5 meter cable, which makes it easier to measure in hard-to-reach places.
In the flow velocity measurement mode, the device memory can save up to 8 values for various measurement points, from which the average measurement value is then calculated. The minimum and maximum indications are also stored in the memory.
RS 232 port enables the connection of the device with a PC in order to save the current indications after installing the software (not included in the set). It is adapted to be mounted on a tripod.
AZ 8911 is powered by 4 x 1.5V alkaline batteries, type AA. It has an automatic power cut-off after 15 minutes. idle and low battery indication. After switching off the energy saving function it can work in a continuous mode, e.g. during long-term measurements.
APPLICATION
For the control and inspection of chimney, ventilation and air supply drafts and for precise measurements in industry. The necessary equipment for facility administrators, chimney sweeps, manufacturers and service technicians of heating and air conditioning devices, it can also be used for precise measurements in industry, for laboratory purposes, for indoor air quality control, etc.
ADVANTAGES
- LCD display with the option of backlighting indications
- 77mm vane probe placed on a 1.5 meter cable
- Highly sensitive sensors for measuring temperature and humidity
- Measurement of air flow speed and efficiency as well as relative humidity
- Determination of the temperature, melting point and evaporation of WET BULB
- DEW POINT dew point temperature measurement
- Ability to change measurement units (metric or British system)
- Useful functions: HOLD, MAX / MIN, AVG
- 8-point average speed measurement function
- Automatic power cut-off after 15 minutes, with option to deactivate
- Battery level indicator
- Possibility of device cooperation with a PC
Product information | |
Rotor diameter | 77 mm |
Display | LCD 26 x 44 mm, backlit |
Display update | Every 1 second |
Clamping screw | Built-in 1/4 inch mounting screw thread (e.g. tripod) |
Sockets, connectors, ports | RS232 for connection to a computer |
Additional features | HOLD - stopping indications on the display MIN, MAX, AVG - measurement mode of the minimum, maximum and average value 8 average velocity points - measurement of average air velocity from 8 measurement points |
Additionally included | Vane probe 4 x 1.5V DC alkaline batteries, type AA Case User manual |
Measurement of air speed | |
Measurement range | 0.6 ~ 32 m/s |
Accuracy of measurement | ±2% of reading ±0.2 m/s * Accuracy may decrease with use. It is recommended to calibrate the device regularly. |
Measurement resolution | 0.1 |
Units of measurement | m/s, ft/m |
Measurement of air flow efficiency | |
Measurement range | 0 ~ 99999 m3/s |
Measurement resolution | 0.1 (range: 0 ~ 9999.9) 1 (range: 10000 ~ 99999) |
Units of measurement | m3/min, ft3/min |
Humidity measurement | |
Response time | 60 seconds (typical) |
Measurement range | 0.1 ~ 99.9% RH |
Accuracy of measurement | ±3% range: 10 ~ 90% RH (at 25°C) ±5% range: 0 ~ 10% RH and 90 ~ 99.9% RH * Accuracy may decrease with use. It is recommended to calibrate the device regularly. |
Measurement resolution | 0.1 |
Temperature measurement | |
Measurement range | -20 ~ 60°C |
Accuracy of measurement | ±0.6°C * Accuracy may decrease with use. It is recommended to calibrate the device regularly. |
Measurement resolution | 0.1 |
Units of measurement | °C / °F |
WET BULB measurement | |
Measurement range | -22 ~ 70.0°C |
Measurement resolution | 0.1 |
DEW POINT measurement | |
Measurement range | -68 ~ 70.0°C |
Measurement resolution | 0.1 |
Power | |
Power source | 4 x 1.5V DC alkaline batteries, type AA |
Energy saving | Auto power off after 15 minutes of inactivity, with the option to deactivate |
Battery discharge indication | Yes |
Working and storage conditions | |
Working conditions | Temperature: 0 ~ 50°C Humidity: up to 80% RH, non-condensing |
Storage conditions | Temperature: -10 ~ 50°C - without batteries Humidity: up to 90% RH non-condensing |
Quality and safety | |
Certificates and approvals | CE mark |